aileike@outlook.com +86 18915184475
ENEN PTPT IDID RURU ARAR
Facebook
X
Youtube
TIKTOK
TELEGRAM
AILEIKE TT&C EQUIPMENT(HUAIAN)CO., LTD.
ENEN PTPT IDID RURU ARAR
  • Home
  • About us
  • Products
    • Pressure series
      • ELECTRODYNAMIC calibration series
      • Pointer pressure gauge
      • Air pressure calibration
      • Hydraulic Calibration
      • Handheld series
      • pressure calibrator
      • Pressure conversion connector
      • Pressure transmitter
    • Digtal pressure gauge
    • Calibration series
      • Temperature calibration
      • 475
    • Liquid level series
      • Radar level gauge
      • Level gauge
      • guided wave radar
      • Tuning fork/RF admittance
      • Magnetic level gauge
    • Temperature series
      • Thermal resistance
      • Bimetal thermometer
      • Thermocouple
    • Flow series
      • Electromagnetic flowmeter
      • Turbine flowmeter
      • Vortex flowmeter
      • Flow meter
      • Ultrasonic flowmeter
    • Petroleum Machinery
    • Display series
      • Digital display series
      • Recording instrument
    • Accessory
  • News
    • Technical articles
    • Industry-news
  • F.A.Q
  • Contact us
Get a Quote
Quick Inquiry

Get a Quote

Fill in the form below and we will get back to you shortly.

Technical articles

  1. Home
  2. News
  3. Technical articles
  4. Operational Procedures for the Calibrati...

Operational Procedures for the Calibration and Verification of Positive and Negative Pressure Measurements in HART Pressure Transmitters

HART pressure transmitters are critical pressure measurement instruments in industrial production, and regular calibration is essential. For this calibration of HART pressure transmitters, three pieces of equipment were utilised: the ALKT511 low-pressure calibration bench, the ALKC400A digital pressure calibrator, and the ALK475TREX handheld programmer (Chinese version). The process was carried out in accordance with established procedures, covering environmental control, wiring, zero and span calibration, pressure ramp-up and ramp-down, and hysteresis verification. This ensures a linear relationship between pressure and the 4–20 mA current signal, thereby eliminating measurement errors. Furthermore, regular verification and maintenance of the equipment are required to lay the foundation for the long-term stable operation of the transmitters, the accuracy of the data, and the proper functioning of the automatic control system.

I. Tools for Calibrating HART Pressure Transmitters

Name

Low-pressure calibration bench

Digital pressure calibrator

HART handheld communicator

Pressure transmitter

Model

ALKT511

ALKC400A

ALK475TREX (Chinese Version)

Various models

Technical Specifications

Pressurisation method: Manual

Pressure: -100 to 100 kPa

Medium: Clean air

Outlets: 2

Connection: M20×1.5(F)

Net weight: 4.6 kg

Product dimensions:

320 × 190 × 140 mm

Pressure: -100 to 100 kPa

Accuracy: ±0.02% FS

Temperature compensation: With temperature compensation

Current measurement: ±25.0000 mA

Voltage measurement: ±50.0000 V

DC output: 24 VDC

Communication Interface: RS-232/RS-485

Power Supply: Built-in rechargeable battery

Connection: M20×1.5 (M)

Net Weight: 1 kg

Dimensions:156×47×239 mm

Communication: HART protocol

Display: Monochrome screen

Language: Chinese

DC Output: 24 V DC

Power Supply: Built-in rechargeable battery

Net Weight: 0.5 kg

Dimensions:230 × 130 × 45 mm

Pressure: -100 to 100 kPa

Accuracy: ±0.075% FS

±0.1% FS

±0.2% FS

±0.5% FS

Output signal: 4–20 mA + HART

Power supply: 24 V DC

Connection: M20×1.5 (M)

Product Dimensions

The entire calibration process must strictly comply with current metrological technical specifications. The primary references include:JJG 882-2019 “Pressure Transmitters,” which applies to the verification, periodic calibration, and in-service inspection of pressure transmitters, and is used to standardize the measurement error, hysteresis, and repeatability of transmitters.

II. Detailed Operational Procedures

(1) Preliminary Preparations

1. Environmental Requirements

Maintain a temperature of (20±2)°C and a relative humidity of ≤85% RH, ensuring there is no condensation. Ensure the site is free of vibration, dust, and corrosive gases.

2. Tool Connection

• Connect the pressure transmitter and the ALKC400A digital pressure calibrator to the ALKT511 outlet using a sealed connection.

• Wiring Method 1 (24V Power Supply for Digital Pressure Calibrator):

(1)Red test lead connection: Insert one end into the standard 24V meter jack, and clip the other end to the transmitter's positive terminal.

(2)Connect a 250-ohm resistor to the negative terminal of the transmitter

(3)Black test lead connection: Insert one end into the standard meter's current terminal, and clip the other end to one side of the 250Ω resistor.

(4)Handheld unit communication cable connection: Plug one end into the handheld unit’s communication port, and attach the alligator clips on the other end to the two terminals of the 250Ω resistor.

(5)Complete the wiring to form a closed circuit

Wiring Diagram (1)

• Wiring Method 2 (24 V power supply for the ALK475TREX handheld unit):

(1)Red test lead connection: Insert one end into the standard meter's current terminal, and clip the other end to the transmitter's negative terminal.

(2)Black test lead connection: Insert one end into the common terminal of the standard meter.

(3)Handheld unit communication cable connection: Plug one end into the handheld unit’s communication port; attach the red clip to the transmitter’s positive terminal, and connect the black clip to the clip on the black test lead.

(4)Complete the wiring to form a closed circuit, then turn on the 24V switching power supply of the ALK475TREX handheld unit.

Wiring Diagram (2)

• The transmitter reads 0 at the current atmospheric pressure; a standard manometer measures the transmitter’s current as 12 mA.

(2) The process of positive pressure calibration for HART pressure transmitters

1. Boost calibration process

• First, open the ALKT511 regulating valve, turn the fine-adjustment knob anticlockwise until it stops, and close the relief valve.

• Use the fine-adjustment knob to increase the pressure clockwise to the calibration value of 50 kPa. If the pressure reading from the transmitter deviates from 50 kPa, or if the current reading of 16 mA from the transmitter as measured by the standard gauge deviates, use the handheld calibrator to calibrate the pressure and current values.

• Calibrating the transmitter’s pressure reading:

Select either the low-range fine-tuning or high-range fine-tuning. If the transmitter’s measured pressure is below 50 kPa, select low-range fine-tuning; if it is above 50 kPa, select high-range fine-tuning. Then press the right button, enter the pressure value of 50 kPa, and finally press the right button again to confirm. Both the pressure and current values will be successfully calibrated.

• Follow the same procedure, using the fine-tuning knob to increase the pressure clockwise until it reaches the calibration value of 100 kPa. The current value should be 20 mA. If there is an error, calibrate using the handheld programmer (as described above).

2. Pressure reduction and calibration process

• Use the fine-tuning knob to turn the pressure downwards to 50 kPa. The current reading on the standard gauge should be 16 mA. If there is any deviation, use the handheld calibrator to adjust the pressure and current values.

• Record the relationship between pressure and current at each point, and verify that the hysteresis (the difference in current at the same point during pressure increase and decrease) meets the requirement (≤0.05% FS).

3. Pressure relief process

Open the pressure relief valve and gradually release the pressure in the pipe; once the pressure has been reduced to zero, the measurement is complete.

(3) Calibration procedure for HART pressure transmitters under negative pressure

1. Boost calibration process

• With the ALKT511 pressure-regulating valve in the open position, turn the fine-adjustment knob clockwise until it stops, then close the relief valve.

• Use the fine-adjustment knob to apply pressure in an anti-clockwise direction until the pressure reaches the calibration value of -50 kPa. If there is a deviation between the pressure value measured by the transmitter and -50 kPa, or if there is a deviation in the 8 mA current value measured by the standard gauge, use the handheld calibrator to calibrate the pressure and current values.

• Calibrating the transmitter’s pressure reading:

Select either the low-range or high-range fine-tuning. If the transmitter’s measured pressure is below -50 kPa, select low-range fine-tuning; if it is above -50 kPa, select high-range fine-tuning. Then press the right button, enter the pressure value -50 kPa, and finally press the right button again to confirm. Both the pressure and current values will be successfully calibrated.

• Follow the same procedure, using the fine-tuning function to apply pressure in an anti-clockwise direction. Apply pressure up to the calibration value of -100 kPa; the current value should be 4 mA. If there is an error, calibrate using the handheld programmer (as described above).

2. Pressure reduction and calibration process

• Use the fine-tuning knob to turn clockwise and reduce the pressure to -50 kPa. The current reading on the standard gauge should be 8 mA. If there is a deviation, use the handheld calibrator to adjust the pressure and current values.

• Record the relationship between pressure and current at each point, and verify that the hysteresis (the difference in current at the same point during pressure increase and decrease) meets the requirement (≤0.05% FS).

3. Pressure relief process

Open the pressure relief valve and gradually release the pressure in the pipe; once the pressure has been reduced to zero, the measurement is complete.

III. Important Notes

Proper use

• Operations must be carried out in accordance with standard procedures. When pressurising or depressurising the ALKT511 micro-pressure calibration bench, do so slowly to avoid pressure surges.

• Ensure the connection between the ALKC400A digital pressure calibrator and the handheld controller is correct; take great care to prevent reverse connection.

• Readings at calibration points must be taken only after the standard gauge reading has stabilised, ensuring accurate and real-time recording.

• Upon completion of the test, first slowly depressurise to zero, ensuring there is no residual pressure, then switch off the power supply before removing the instrument under test.

Maintenance and calibration intervals

• Perform metrological verification of the ALKC400A digital pressure calibrator at regular intervals (recommended once a year) to ensure measurement accuracy meets the required standards.

• Switch off the power supply when the pressure gauge is not in use; if it is not used for an extended period, it should be charged once every three months.

• If the micro-pressure stand is not to be used for an extended period, it should be depressurised before storage and covered with a cloth cover to prevent dust from entering the instrument.

IV. Conclusion

This project involved the full-process calibration of HART pressure transmitters for both positive and negative pressures. Utilising the coordinated operation of the ALKT511, ALKC400A and ALK475TREX, the team accurately completed the wiring, zero-point adjustment, span calibration and verification. By manually correcting the deviation between the pressure reading and the 4–20 mA current signal using a HART handheld communicator, instrument performance was optimised to ensure the equipment met performance standards. Adhering to standard operating procedures mitigates risks, enhances measurement stability, and strengthens calibration and maintenance protocols, thereby ensuring the long-term reliable operation of the instruments.

Related News

Recommended Pressure Standards - ALKB201

Recommended Pressure Standards - ALKB201

Recommended Pressure Standards - ALKB9601

Recommended Pressure Standards - ALKB9601

Recommended Pressure Standards - ALKB202

Recommended Pressure Standards - ALKB202

Product Categories

  • Pressure series
    • ELECTRODYNAMIC calibration series
    • Pointer pressure gauge
    • Air pressure calibration
    • Hydraulic Calibration
    • Handheld series
    • pressure calibrator
    • Pressure conversion connector
    • Pressure transmitter
  • Digtal pressure gauge
  • Calibration series
    • Temperature calibration
    • 475
  • Liquid level series
    • Radar level gauge
    • guided wave radar
    • Level gauge
    • Tuning fork/RF admittance
    • Magnetic level gauge
  • Temperature series
    • Thermal resistance
    • Bimetal thermometer
    • Thermocouple
  • Flow series
    • Electromagnetic flowmeter
    • Turbine flowmeter
    • Vortex flowmeter
    • Flow meter
    • Ultrasonic flowmeter
  • Petroleum Machinery
  • Display series
    • Digital display series
    • Recording instrument
  • Accessory

Links

Google.comBaidu.com

侧边1

侧边2

Latest Post

  • Recommended Pressure Standards - ALKB9601
    Recommended Pressure Standards - ALKB960
    2026-06-12
  • Operational Procedures for the Calibration and Verification of Positive and Negative Pressure Measurements in HART Pressure Transmitters
    Operational Procedures for the Calibrati
    2026-06-15
  • Recommended Pressure Standards - ALKB202
    Recommended Pressure Standards - ALKB202
    2026-06-11

Feedback

Contact Us

3841141734

+86 18915184475

+86-18915184475

aileike@outlook.com

No. 6, Yinjizhen Industrial Park, Jiangsu Province

Mobile Website

Mobile QR Code


Copyright © 2013-2020 AILEIKE TT&C EQUIPMENT(HUAIAN)CO., LTD. All Rights Reserved.

https://www.youtube.com/@aileike https://x.com/AiLeike1201 https://vk.ru/id1085515017 https://www.youtube.com/@aileike https://www.tiktok.com/@aileike01 +86 18915184475 aileike@outlook.com +86-18915184475
WeChat
WeChat QR Code