aileike@outlook.com +86 18915184475
ENEN PTPT IDID RURU ARAR
Facebook
X
Youtube
TIKTOK
TELEGRAM
AILEIKE TT&C EQUIPMENT(HUAIAN)CO., LTD.
ENEN PTPT IDID RURU ARAR
  • Home
  • About us
  • Products
    • Pressure series
      • ELECTRODYNAMIC calibration series
      • Pointer pressure gauge
      • Air pressure calibration
      • Hydraulic Calibration
      • Handheld series
      • pressure calibrator
      • Pressure conversion connector
      • Pressure transmitter
    • Digtal pressure gauge
    • Calibration series
      • Temperature calibration
      • 475
    • Liquid level series
      • Radar level gauge
      • Level gauge
      • guided wave radar
      • Tuning fork/RF admittance
      • Magnetic level gauge
    • Temperature series
      • Thermal resistance
      • Bimetal thermometer
      • Thermocouple
    • Flow series
      • Electromagnetic flowmeter
      • Turbine flowmeter
      • Vortex flowmeter
      • Flow meter
      • Ultrasonic flowmeter
    • Petroleum Machinery
    • Display series
      • Digital display series
      • Recording instrument
    • Accessory
  • News
    • Technical articles
    • Industry-news
  • F.A.Q
  • Contact us
Get a Quote
Quick Inquiry

Get a Quote

Fill in the form below and we will get back to you shortly.

Technical articles

  1. Home
  2. News
  3. Technical articles
  4. A Comprehensive Analysis of the Entire P...

A Comprehensive Analysis of the Entire Process for Verifying the Opening and Closing of Pressure Switch Contacts

Pressure switches are core measurement and control components in industrial hydraulic systems, automatically opening and closing circuits in response to changes in pressure. Deviations in pressure switch parameters can easily lead to inaccurate calibration readings; therefore, it is essential to carry out regular pressure measurements, performance tests and set-point calibration. This calibration was carried out under standard environmental conditions of (20±2) °C and relative humidity ≤ 85% RH, using an ALKT803 hydraulic calibration bench and an ALKC400A digital pressure calibrator to test the operational status of the pressure switch’s normally open and normally closed contacts respectively. By applying and releasing pressure gradually to test the contact make-and-break behaviour, operational risks such as pressure surges and wiring errors were avoided, ensuring the authenticity and accuracy of the test data.

I. Tools for Calibrating Pressure Switches

Name

Hydraulic calibration bench

Digital Pressure Calibrator

Pressure switch

Model

ALKT803

ALKC400A

Various models

Technical Specifications

Pressurisation method: Manual

Pressure range: 0–60 MPa

Medium: Purified water or calibration oil

Outlets: 3

Connection: M20×1.5 (F)

Net weight: 14 kg

Product dimensions: 410×410×180 mm

 

Pressure: 0–60 MPa

Accuracy: 0.02% FS

Temperature compensation: With temperature compensation

Current measurement: ±25.0000 mA

Voltage measurement: ±50.0000 V

DC output: 24 V DC

Communication interface: RS-232/RS-485

Power supply: Built-in rechargeable battery

Connection: M20×1.5 (M)

Net weight: 1 kg

Dimensions: 156 × 47 × 239 mm

 

Pressure switch: 0–60 MPa

Set point: 30 MPa

Contact type: Normally open/normally closed switch

Wetted parts material: Stainless steel

Mounting thread: M20×1.5 (default)

 

Product Dimensions

 

 

 

The entire verification process must be carried out in strict accordance with current metrological technical specifications. The main references include:JJG 544-2011 ‘Pressure Controllers’, which applies to the initial verification, subsequent verification and in-service inspection of pressure, vacuum and pressure-vacuum controllers (including mechanical and electronic types); these are pressure control devices that operate without an external power supply and rely on pressure changes to directly drive mechanical components to close or open contacts.

II. Detailed Explanation of the Operational Process

(1) Preliminary Preparations

1. Environmental Requirements

The temperature must be maintained at (20±2) °C, with relative humidity ≤85% RH and no condensation. Ensure that the site is free from vibration, dust and corrosive gases.

2. Filling, Pressurising and Press-through of the Hydraulic Test Bench

• Prepare the pressure-transmitting medium (purified water or calibration oil).

• First, unscrew the pressure relief valve on the ALKT803 hydraulic calibration bench and fill the reservoir with the medium, ensuring it is sufficiently filled (to at least two-thirds capacity). Once filling is complete, reattach and tighten the pressure relief valve.

• Next, purge any excess air from the hydraulic bench. First, unscrew the plugs from the three outlet ports, open the pressure-regulating valve, relief valve and overpressure shut-off valve, turn the fine-adjustment valve anti-clockwise to the appropriate position, and close the relief valve. Then, use the lever to pressurise the system, directing the pressure to the three outlet ports.

3. Connecting the equipment

• Connect the pressure switch and the ALKC400A digital pressure calibrator to the outlet of the ALKT803 hydraulic calibration bench using sealed connections. Press the Switch/Peak button on the ALKC400A to switch to pressure switch measurement mode.

• Pressure switch pre-adjustment: Open the protective cover of the adjustment knob on top of the pressure switch, have the adjustment tool ready, and allow sufficient travel for clockwise (increase) and anti-clockwise (decrease) rotation to set the pressure switch setpoint.

• Wiring instructions:

(1) Connect the red test lead of the ALKC400A to the 24V output terminal of the standard gauge, and the black test lead to the pressure switch measurement terminal.

(2) Clip the other end of the red alligator clip to the common terminal (COM) of the pressure switch, and the black alligator clip to the normally open (NC) or normally closed (NO) terminal.

(3) Now, first clip the black crocodile clip to the normally open (NC) terminal. At this point, the ALKC400A pressure switch display will show the word ‘Open’, confirming that the pressure switch’s normally open (open circuit) state is functioning correctly and that the wiring is correct.

 

(2) Procedure

Key calibration steps: Set the target pressure to 30 MPa. Rotate the pressure switch adjustment knob slightly (clockwise to increase pressure, anti-clockwise to decrease pressure), and carry out pressure testing using the calibration bench to precisely calibrate the pressure switch’s actuation point to the target pressure of 30 MPa. Once calibration is complete, keep the knob in its set position.

1. Testing the closing and opening of the pressure switch whilst in the normally open state

• Apply pressure using the ALKT803 lever; stop when the pressure reaches approximately 5–10 MPa. Close the pressure-regulating valve and the overpressure shut-off valve, then use the fine-adjustment knob to turn clockwise until the upper cut-off value (actuation pressure) for closing the switch reaches 30 MPa. You will hear a ‘beep’ from the instrument’s buzzer, the contact status will change, and the switch test on the standard gauge interface will change from ‘open’ to ‘closed’. the pressure switch switches from the open state to the closed state; record the actuation pressure value.

• Continue to reduce the pressure by turning the fine-adjustment knob anti-clockwise until the lower cut-off value (reset value) of 27 MPa is reached; again, a “beep” from the instrument will sound, the contact status will change, and the switch test on the standard gauge interface will change from “closed” to “open”; the pressure switch switches from the closed state to the open state; record the reset pressure value.

• Once the test is complete, open the pressure relief valve, pressure-regulating valve and overpressure shut-off valve in sequence to reduce the pressure to 0; the measurement is then complete.

2. Testing the closing and opening of a pressure switch in the normally closed state

 

• First, clip the black alligator clip onto the normally open (NO) terminal; you will then see that the pressure switch is in the normally closed state, and the ALKC400A pressure switch test display will show the word ‘Open’.

• Open the pressure-regulating valve, the relief valve and the overpressure shut-off valve; turn the fine-adjustment valve anti-clockwise to the appropriate position, then close the relief valve.

• Apply pressure using the lever, stopping at approximately 5–10 MPa. Close the pressure-regulating valve and the overpressure shut-off valve, then use the fine-adjustment valve to turn clockwise to the upper cut-off value (actuation value) of 30 MPa at which the switch opens. You will hear a ‘beep’ from the instrument’s buzzer, the contact status will change, and the switch test on the standard gauge interface will change from ‘Open’ to ‘Closed’. the pressure switch switches from the closed state to the open state; record the actuation pressure value.

• Continue to reduce the pressure by turning the fine-adjustment valve anti-clockwise until it reaches the lower cut-off value (reset value) of 27 MPa; again, you will hear a ‘beep’ from the instrument, the contact status will change, the ‘Open/Closed’ indicator on the standard gauge will change from ‘Open’ to ‘Closed’, and the pressure switch will switch from the open state to the closed state; record the reset pressure value.

• Once the test is complete, open the pressure relief valve, pressure-regulating valve and overpressure shut-off valve in sequence to reduce the pressure to 0; the measurement is then complete.

III. Points to Note

Proper Use

• Operations must be carried out in accordance with standard procedures. Pressure must be increased and decreased gradually on the ALKT803 hydraulic calibration stand to avoid pressure surges.

• The wiring of the ALKC400A digital pressure calibrator must be clearly identified; take great care to prevent reverse connections.

• Readings at calibration points must be taken only after the value on the standard gauge has stabilised, ensuring accurate and real-time recording.

• Upon completion of the test, first slowly depressurise the system to zero; the instrument under test may only be removed once all residual pressure has been released.

Maintenance and Calibration Intervals

• The ALKC400A digital pressure calibrator must undergo periodic metrological verification (recommended once a year) to ensure measurement accuracy meets standards.

• Switch off the power supply when the pressure gauge is not in use; if not used for an extended period, recharge the battery once every three months.

• If the hydraulic test bench is not to be used for an extended period, depressurise it before storage, empty the fluid from the reservoir cup and cover it with a cloth cover to prevent dust from entering the instrument.

IV. Conclusion

Pressure measurement, contact testing and set-point calibration of pressure switches are key to ensuring the reliable operation of hydraulic systems. In this instance, based on on-site operating conditions, we conducted item-by-item testing of both normally open and normally closed switch configurations, precisely measuring the critical pressures for activation and deactivation, verifying actuation and reset response performance, and identifying potential hazards such as contact failure and set-point drift. Standardised calibration effectively corrects parameter deviations, ensuring the switches operate reliably at the set pressure. A system of periodic testing should be established for routine calibration and measurement. Pressure switches should be retested and calibrated at regular intervals to promptly eliminate performance deviations, minimise the risk of overpressure failures, and effectively ensure the long-term, stable operation of the entire hydraulic system.

 

Related News

Recommended Pressure Standards - ALKB201

Recommended Pressure Standards - ALKB201

AILEIKE 100 MPa High-Pressure Switch Calibration Procedure and Metrological Testing Specifications

AILEIKE 100 MPa High-Pressure Switch Calibration Procedure a

Recommended Pressure Standards - ALKB9601

Recommended Pressure Standards - ALKB9601

Practical Guide to the Calibration of Positive and Negative Pressure Measurements for Differential Pressure Transmitters

Practical Guide to the Calibration of Positive and Negative

Operational Guide to the Precise Calibration and Comprehensive Testing of Pressure Switches

Operational Guide to the Precise Calibration and Comprehensi

AILEIKE Operational Guidelines for the Calibration of Stainless Steel Pressure Gauges: Positive and Negative Pressure, Full Range

AILEIKE Operational Guidelines for the Calibration of Stainl

Product Categories

  • Pressure series
    • ELECTRODYNAMIC calibration series
    • Pointer pressure gauge
    • Air pressure calibration
    • Hydraulic Calibration
    • Handheld series
    • pressure calibrator
    • Pressure conversion connector
    • Pressure transmitter
  • Digtal pressure gauge
  • Calibration series
    • Temperature calibration
    • 475
  • Liquid level series
    • Radar level gauge
    • guided wave radar
    • Level gauge
    • Tuning fork/RF admittance
    • Magnetic level gauge
  • Temperature series
    • Thermal resistance
    • Bimetal thermometer
    • Thermocouple
  • Flow series
    • Electromagnetic flowmeter
    • Turbine flowmeter
    • Vortex flowmeter
    • Flow meter
    • Ultrasonic flowmeter
  • Petroleum Machinery
  • Display series
    • Digital display series
    • Recording instrument
  • Accessory

Links

Google.comBaidu.com

侧边1

侧边2

Latest Post

  • A Comprehensive Operational Guide to the Full Process of Accurate Calibration, Testing and Measurement of Pressure Gauges
    A Comprehensive Operational Guide to the
    2026-06-24
  • Recommended Pressure Standards - ALKB9601
    Recommended Pressure Standards - ALKB960
    2026-06-12
  • AILEIKE 100 MPa High-Pressure Switch Calibration Procedure and Metrological Testing Specifications
    AILEIKE 100 MPa High-Pressure Switch Cal
    2026-07-11

Feedback

Contact Us

3841141734

+86 18915184475

+86-18915184475

aileike@outlook.com

No. 6, Yinjizhen Industrial Park, Jiangsu Province

Mobile Website

Mobile QR Code


Copyright © 2013-2020 AILEIKE TT&C EQUIPMENT(HUAIAN)CO., LTD. All Rights Reserved.

https://www.youtube.com/@aileike https://x.com/AiLeike1201 https://vk.ru/id1085515017 https://www.youtube.com/@aileike https://www.tiktok.com/@aileike01 +86 18915184475 aileike@outlook.com +86-18915184475
WeChat
WeChat QR Code