Standardized metrological verification and calibration procedure for stainless steel pressure gauges
In industrial production, the measurement accuracy of pressure gauges is crucial to the stability of process operation, necessitating regular standardized calibration. This calibration strictly adheres to metrological specifications, utilizing the ALKT804 hydraulic calibration bench and ALKC400A digital pressure calibrator to verify the Y-100B-F stainless steel pressure gauge. The calibration must be conducted in an environment with satisfactory temperature and humidity levels and free from interference. After completing preparations such as medium filling, pipeline exhaust, and equipment connection, the calibration follows a step-by-step process of pressure increase, stabilization, and decrease. After stabilizing the pressure at each verification point, the gauge case is lightly tapped to read the value and calculate the error. Throughout the entire process, pressure is adjusted slowly to prevent pressure shock. Furthermore, regular calibration of standard equipment and maintenance of the hydraulic system ensure long-term stable operation of the equipment.
1. Calibration tool for pressure gauge

The entire verification process must strictly adhere to the current metrological technical specifications, primarily based on JJG52-2013 "Elastic Element Type General Pressure Gauges, Pressure Vacuum Gauges, and Vacuum Gauges". This standard is applicable to the initial verification, subsequent verification, and in-service inspection of elastic element type general pressure gauges, pressure vacuum gauges, and vacuum gauges, covering locally indicating pressure measuring instruments commonly used in industry.
II. Detailed Explanation of Operational Procedures
(I) Preliminary preparations
1. Environmental requirements
The temperature should be controlled at (20±2)℃, with a relative humidity of ≤85%RH, and there should be no condensation. Ensure that the site is free from vibration, dust, and corrosive gases.
2. Hydraulic platform liquid injection, pressurization, conduction, and tool connection
• Prepare the pressure transmission medium (pure water or calibration oil).
First, unscrew the pressure relief valve of the ALKT804 hydraulic calibration bench, inject the medium into the reservoir cup, ensuring that the medium is injected in sufficient quantity (no less than two-thirds). After the injection is complete, reinstall and tighten the pressure relief valve.
• Next, discharge the excess gas from the hydraulic platform. First, unscrew the plugs from the three output ports, open the pressure regulating valve, pressure relief valve, and overpressure cutoff valve, adjust the fine adjustment valve counterclockwise to the appropriate position, and close the pressure relief valve. Then, use a lever to apply pressure, allowing the pilot pressure to reach the three output ports. After the excess gas is discharged, seal and connect the standard gauge ALKC400A and the verified pressure gauge Y-100BF to the ALKT804 output port.
(II) Pressure gauge calibration process
1. Voltage boosting process
• Use the ALKT804 lever to apply pressure, stopping at around 10~15MPa. Close the pressure regulating valve and the overpressure cutoff valve, then use fine adjustment to increase the pressure clockwise to the verification point of 20MPa. After the pressure stabilizes and the standard indicated value no longer fluctuates, check the pressure value of the verified pressure gauge. Then, gently tap the side of the verified pressure gauge's housing with your finger (do not tap the glass or tap hard), observe the pointer position, and record the indicated value after the tapping stabilizes. Use the indicated value after tapping as the actual measured verification reading for this pressure point, and calculate the error.
• Use the trim valve to adjust clockwise to a pressure value of 40 MPa. Once the pressure stabilizes, tap the pressure gauge being tested and record the value.
• For the same operation, use the trim valve to increase the pressure clockwise to pressure values of 60MPa, 80MPa, and 100MPa, check the pressure values on the pressure gauge, and gently tap the verified pressure gauge. Record the indicated value after the tap becomes stable.
2. Decompression process
• Directly use the fine adjustment to decrease the pressure counterclockwise, with pressure values of 80MPa, 60MPa, 40MPa, and 20MPa. Check the pressure values on the pressure gauge, and gently tap the verified pressure gauge. Record the indicated value after the tap stabilizes.
3. Pressure relief process
Open the pressure relief valve, pressure regulating valve, and overpressure cutoff valve in sequence, set the pressure relief to 0, and the measurement is complete.
III. Precautions
Standardized use
• During verification, it is important to ensure that the allowable error of the standard instrument is smaller than that of the instrument being verified, in order to guarantee the accuracy and reliability of the verification results and comply with metrological specifications.
• The pressure transmission medium should be clean and free of impurities to prevent blockage of pipelines and damage to the hydraulic platform and pressure gauge.
• Operations should follow standard procedures. The ALKT804 hydraulic calibration bench should be pressurized and depressurized slowly to avoid pressure shock.
• Increase the pressure to the verification point and record the indicated value after the standard gauge stabilizes. For the verified pressure gauge, gently tap the gauge case after the pressure stabilizes, and take the indicated value after tapping as the accurate reading to ensure accurate recording.
• After the inspection is completed, slowly release the pressure to zero and ensure there is no residual pressure before removing the inspected instrument.
Maintenance and calibration cycle
• Regularly conduct metrological verification on the digital pressure calibrator ALKC400A (it is recommended to do so once a year) to ensure that the measurement accuracy meets the standard.
• When the pressure gauge is not in use, the power should be turned off. If it is not used for a long time, it should be charged once every three months.
• If the hydraulic bench is not to be used for an extended period, it needs to be depressurized before storage. Pour out the liquid in the reservoir cup and cover it with a cloth cover to prevent dust from entering the instrument.
IV. Conclusion
The entire process of this pressure gauge verification strictly adheres to metrological operation guidelines, rigorously implements various standard operating procedures and reading specifications, effectively avoids common detection issues such as pressure shock, reading deviation, and pipeline impurity interference, and ensures that the measurement and verification data are objective and effective. In conjunction with the maintenance requirements of supporting equipment, control measures such as annual verification of standard instruments and idle maintenance of hydraulic devices are implemented. As a core instrument for industrial pressure monitoring, regular and compliant verification of pressure gauges can ensure the accuracy of verification values, standardize metrological testing work, and lay a solid metrological foundation for the smooth operation of industrial production.