Practical Guide for Calibration of AILEIKE 60MPa Dial Pressure Gauge
The measurement accuracy of the pointer pressure gauge directly affects the adjustment of production process parameters, and periodic verification is a crucial step in maintaining the integrity of the measurement traceability. For this verification, the ALKT9803 hydraulic calibration bench and ALKC400D precision digital pressure gauge were selected to conduct metrological verification on the Y-100BF pointer pressure gauge. The verification was carried out in a standard environment with controllable temperature and humidity, free from impact vibration and corrosive media. Standardized preliminary operations such as liquid injection, exhaust, and instrument connection were performed. According to the verification standards, multi-point pressure rise and stabilization, light tapping on the gauge case to read values, pressure reduction for re-verification, and pressure relief operations were completed. Inspection data was collected, and operation specifications and regular maintenance plans for the verification equipment were formulated to ensure the reliability of the pressure gauge readings and meet the equipment pressure monitoring and metrological compliance requirements.
1. Calibration tool for pressure gauge

The entire verification process must strictly adhere to the current metrological technical specifications, primarily based on JJG52-2013 "General Pressure Gauges, Pressure Vacuum Gauges, and Vacuum Gauges with Elastic Elements". This standard is applicable to the initial verification, subsequent verification, and in-service inspection of general pressure gauges, pressure vacuum gauges, and vacuum gauges with elastic elements, covering locally indicating pressure measuring instruments commonly used in industry.
II. Detailed Explanation of Operational Procedures
(I) Preliminary preparations
1. Environmental requirements
The temperature should be controlled at (20±2)℃, with a relative humidity of ≤85%RH, and there should be no condensation. Ensure that the site is free from vibration, dust, and corrosive gases.
2. Hydraulic platform liquid injection, pressurization, conduction, and tool connection
• Prepare the pressure transmission medium (pure water or calibration oil) and a syringe.
First, unscrew the exhaust knob on the ALKT9803 storage cup. Use a syringe to draw up the medium and inject it into the storage cup. Ensure that enough medium is injected (no less than two-thirds of the cup). After the injection is complete, tighten the knob securely.
• Next, discharge the excess gas from the hydraulic bench. First, unscrew the plugs of the two output ports, then loosen the knob on the liquid injection port, and open the stop valve (just loosen it half a turn). Adjust the micro-adjustment knob counterclockwise until it stops, and then close the pressure relief valve. Next, use the hand pressure lever to apply pressure, allowing the pressure guiding liquid to flow to the two output ports. After the excess gas is discharged, connect the standard gauge ALKC400D and the verified pressure gauge Y-100BF to the ALKT9803 output port in a sealed manner, and then close the stop valve.
(II) Pressure gauge calibration process
1. Voltage boosting process
• Use the hand pressure lever to gradually increase the pressure, stopping when it reaches approximately 5~10MPa. Then, use fine adjustment to increase the pressure clockwise to the verification point of 10MPa. Once the pressure stabilizes and the standard indication no longer fluctuates, check the pressure value of the verified pressure gauge. Next, gently tap the side of the verified pressure gauge's casing with your finger (tapping on the glass or hitting hard is strictly prohibited), observe the pointer position, and record the indication after the tapping stabilizes. Use the indication after tapping as the actual measured verification reading for this pressure point, and calculate the error.
• Then, use fine adjustment to increase the pressure clockwise to the verification point of 20 MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the value.
• For the same operation, directly use the fine adjustment to pressurize clockwise to the verification points of 30MPa, 40MPa, 50MPa, and 60MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the values.
2. Decompression process
Directly use the fine adjustment to decompress counterclockwise, reducing the pressure to the verification points of 50MPa, 40MPa, 30MPa, 20MPa, and 10MPa. Check the pressure value of the verified pressure gauge, tap the verified pressure gauge, and record the indicated value after the tap stabilizes.
3. Pressure relief process
Open the pressure relief valve and the stop valve in sequence, gradually releasing the pressure inside the pipeline until it reaches 0. The measurement is complete.
III. Precautions
Standardized use
• The pressure transmission medium needs to be clean and free of impurities to prevent blockage of pipelines, damage to the hydraulic platform, and pressure gauge.
• Operations should follow standard procedures. The ALKT9803 hydraulic calibration bench should be pressurized and depressurized slowly to avoid pressure shock.
• Increase the pressure to the verification point and record the indicated value after the standard gauge stabilizes. For the verified pressure gauge, gently tap the gauge case after the pressure stabilizes, and take the indicated value after tapping as the accurate reading to ensure accurate recording.
• After the inspection, slowly release the pressure to zero first, ensuring there is no residual pressure, before removing the inspected instrument.
Maintenance and calibration cycle
• Regularly conduct metrological verification on the Precision digital pressure gauge alkc400D (it is recommended to do so once a year) to ensure that the measurement accuracy meets the standard.
• When the pressure gauge is not in use, the power should be turned off. If it is not used for a long time, it should be charged once every three months.
• If the hydraulic bench is not to be used for an extended period, it should be depressurized before storage. Pour out the liquid in the reservoir cup and cover it with a cloth cover to prevent dust from entering the instrument.
IV. Conclusion
This time, we used the ALKT9803 hydraulic calibration bench from AILEIKE, paired with the ALKC400D precision digital pressure gauge, to conduct a full-range calibration test for the Y-100BF pointer pressure gauge, covering a range of 0~60MPa. The operation was conducted to avoid the impact of vibration and corrosive environments, and standardized procedures were followed for medium filling, pipeline exhaust, and instrument connection. We performed point-by-point pressure increase and stabilization, return pressure decrease and reading, error evaluation, and troubleshooting of instrument return errors. Standardized operation and instrument maintenance management ensured the authenticity and integrity of the calibration data, providing a basis for pressure gauge value transfer and compliance monitoring of small and medium-sized pressure-bearing devices.