How to calibrate a pressure gauge: Operating specifications for full-range calibration of positive pressure and vacuum
Pressure gauges are essential measurement and control devices in industrial pressure-bearing systems, and their indication accuracy is a crucial foundation for ensuring production and process compliance. In this article, the ALKT9603 air pressure calibration bench and ALKC400E digital pressure calibrator are used to calibrate and verify the Y-100BF stainless steel pressure gauge. Strict control is maintained over the operating environmental conditions, with two major working conditions: positive pressure and vacuum. The standardized process of pressure rise and stabilization, light tapping for reading, pressure drop and re-verification, and pressure relief is followed. Data is accurately recorded, measurement errors are calculated, and operational guidelines and equipment maintenance requirements are standardized to ensure the accuracy and reliability of the pressure gauge readings.
1. Calibration tool for pressure gauge

The entire verification process must strictly adhere to the current metrological technical specifications, primarily based on JJG52-2013 "Elastic Element Type General Pressure Gauges, Pressure Vacuum Gauges, and Vacuum Gauges". This standard is applicable to the initial verification, subsequent verification, and in-service inspection of elastic element type general pressure gauges, pressure vacuum gauges, and vacuum gauges, covering locally indicating pressure measuring instruments commonly used in industry.
II. Detailed Explanation of Operational Procedures
(I) Preparatory work
1. Environmental requirements
The temperature should be controlled at (20±2)℃, with a relative humidity of ≤85%RH, and there should be no condensation. Ensure that the site is free from vibration, dust, and corrosive gases.
2. Tool connection
• Connect the standard gauge ALKC400E and the verified pressure gauge Y-100BF in a sealed manner to the ALKT9603 output port.
(II) Pressure gauge calibration process
1. Voltage boosting process
First, open the stop valve of ALKT9603, then push the positive and negative pressure switch valve inwards (do not switch under pressure), and finally, turn the micro-adjustment knob counterclockwise until it cannot be turned any further, and close the pressure relief valve.
• Use the hand pressure lever to apply pressure, stopping when the pressure reaches the verification point. Then, use fine adjustment to slowly turn clockwise to reach the verification point of 1 MPa. After the pressure stabilizes and the standard indicated value no longer fluctuates, check the pressure value of the verified pressure gauge. Next, gently tap the side of the verified pressure gauge's casing with your finger (tapping on the glass or hitting hard is strictly prohibited), observe the pointer position, and record the indicated value after the tap stabilizes. Use the indicated value after the tap as the actual measured verification reading for this pressure point, and calculate the error.
• Then, use the hand pressure lever to apply pressure to approximately 1.9 MPa and 2.9 MPa, and then stop. Use the fine adjustment to slowly turn clockwise to adjust to the verification points of 2 MPa and 3 MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the value.
Before adjusting to the next verification point, first close the stop valve, then use the fine adjustment to slowly turn clockwise to reach the verification point of 4MPa. Once the pressure stabilizes, tap the pressure gauge being verified and record the value.
• Finally, directly use the fine adjustment to slowly turn clockwise to the verification points of 5MPa and 6MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the value.
2. Decompression process
• First, slowly open the pressure relief valve and reduce the pressure until it stops at the verification point. Close the pressure relief valve, and use the fine adjustment valve to slowly adjust it counterclockwise to the verification point of 5 MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the value.
• Perform the same operation at the verification points of 4MPa, 3MPa, 2MPa, and 1MPa, check the pressure values of the verified pressure gauge, tap the verified pressure gauge, and record the indicated value after it stabilizes.
3. Pressure relief process
Open the pressure relief valve and the stop valve in sequence, gradually releasing the pressure inside the pipeline until it reaches 0. The measurement is complete.
(III) Pressure gauge vacuum calibration process
1. Voltage boosting process
First, remove the verified gauge for measuring positive pressure and install the verified gauge for measuring negative pressure. Pull out the positive and negative pressure switch valve (do not switch under pressure). Next, open the stop valve, then turn the ALKT9603 fine adjustment knob clockwise until it stops, and close the pressure relief valve.
• Use the hand pressure lever to evacuate the vacuum until it reaches the verification value, then use the fine adjustment to slowly adjust counterclockwise to the verification point of -0.02MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the value.
• Perform the same operation at the verification points of -0.04MPa, -0.06MPa, -0.08MPa, and -0.095MPa, check the pressure values of the verified pressure gauge, tap the verified pressure gauge, and record the indicated value after it stabilizes.
2. Decompression process
• First, slowly open the pressure relief valve to reduce the pressure to around the verification value and then stop. Close the pressure relief valve, use the fine adjustment valve to slowly adjust clockwise to the verification point of -0.08 MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the value.
• Perform the same operation with the verification points adjusted to -0.06MPa, -0.04MPa, and -0.02MPa, check the pressure values of the verified pressure gauge, tap the verified pressure gauge, and record the indicated value after the tap becomes stable.
3. Pressure relief process
Open the pressure relief valve, gradually release the pressure inside the pipeline, and when the pressure relief reaches 0, the measurement is complete.
III. Precautions
Standardized use
• Operations should follow standard procedures. The ALKT9603 air pressure calibration station should be pressurized and depressurized slowly to avoid pressure shock.
• Increase the pressure to the verification point and record the indicated value after the standard gauge value stabilizes. For the verified pressure gauge, gently tap the gauge case after the pressure stabilizes, and take the indicated value after tapping as the accurate reading to ensure accurate recording.
• After the inspection is completed, slowly release the pressure to zero and ensure there is no residual pressure before removing the instrument being inspected.
Maintenance and calibration cycle
• Regularly conduct metrological verification on the digital pressure calibrator ALKC400E (it is recommended to do so once a year) to ensure that the measurement accuracy meets the required standards.
• When the pressure gauge is not in use, the power should be turned off. If it is not used for a long time, it should be charged once every three months.
• If the air pressure table is not used for a long time, it should be depressurized and stored, covered with a cloth cover to prevent dust from entering the instrument.
IV. Conclusion
This article systematically standardizes the calibration and testing process for stainless steel pressure gauges across the full range of positive pressure and vacuum, strictly controls the operating environmental conditions, and adheres to operational specifications such as pressure switching, pressure increasing and decreasing, and light tapping for reading. By collecting data point by point and calculating indication errors, the metrological performance of the pressure gauge is accurately evaluated. At the same time, it clarifies the instrument operation procedures, annual verification cycle, and daily maintenance and storage standards. Regular implementation of pressure gauge calibration and measurement work can stabilize the metrological accuracy of the instrument, providing reliable metrological support for process control and pressure monitoring.