Specification for calibrating pointer pressure gauge on ALKT9801 hydraulic calibration bench
As a pressure-bearing equipment measurement and monitoring instrument, the accuracy of measurement of the pointer pressure gauge is crucial for process stability, and periodic verification is a necessary step to ensure the traceability of measurement values. For the verification task of the 0~60MPa range, the Aileco ALKT9801 hydraulic calibration bench, paired with the ALKC400D precision digital pressure gauge, was selected to conduct metrological verification on the Y-100BF pointer pressure gauge. The verification environment was strictly controlled for temperature and humidity, and measures were taken to avoid shock and vibration. Standardized operations were carried out, including joint connection, pressure medium filling, and pipeline exhaust. Strictly following the verification standards, multi-point pressure increase operations were conducted, the gauge case was tapped lightly to read the indicated value, the pressure was reduced for re-verification, and the pressure was released at a constant speed. Recordings of indicated value errors were made, and standardized operation specifications and maintenance plans were formulated to ensure accurate pressure measurement, pressure monitoring, and compliance with measurement requirements.
1. Calibration tool for pressure gauge

The entire verification process must strictly adhere to the current metrological technical specifications, primarily based on JJG52-2013 "Elastic Element Type General Pressure Gauges, Pressure Vacuum Gauges, and Vacuum Gauges". This standard is applicable to the initial verification, subsequent verification, and in-service inspection of elastic element type general pressure gauges, pressure vacuum gauges, and vacuum gauges, covering locally indicating pressure measuring instruments commonly used in industry.
II. Detailed Explanation of Operational Procedures
(I) Preliminary preparations
1. Environmental requirements
The temperature should be controlled at (20±2)℃, with a relative humidity of ≤85%RH, and there should be no condensation. Ensure that the site is free from vibration, dust, and corrosive gases.
2. Hydraulic platform liquid injection, pressurization, conduction, and tool connection
• Prepare the pressure transmission medium (pure water or calibration oil) and a syringe.
First, unscrew the exhaust knob on the ALKT9801 liquid storage cup. Use a syringe to draw up the medium and inject it into the storage cup. Ensure that enough medium is injected (no less than two-thirds of the cup). After the injection is complete, tighten the knob securely.
• Next, discharge the excess gas from the hydraulic bench. First, unscrew the plugs of the two output ports, then loosen the knob on the liquid injection port, and open the output stop valve (just loosen it half a turn). Turn the pressure regulating handle counterclockwise until it stops, and close the pressure relief valve. Then, use the pressure regulating handle to apply pressure, allowing the pilot pressure to reach the two output ports. After the excess gas is discharged, seal and connect the standard gauge ALKC400D and the verified pressure gauge Y-100BF to the ALKT9801 output port.
(II) Pressure gauge calibration process
1. Voltage boosting process
• Use the pressure regulating handle to apply pressure clockwise to the verification point of 10MPa. After the pressure stabilizes and the standard indicated value no longer fluctuates, check the pressure value of the verified pressure gauge. Then, gently tap the side of the verified pressure gauge's housing with your finger (tapping on the glass or hitting hard is strictly prohibited), observe the position of the pointer, and record the indicated value after the tapping stabilizes. Use the indicated value after tapping as the actual measured verification reading for this pressure point, and calculate the error.
• Then, use the pressure regulating handle to increase the pressure clockwise to the verification point of 20 MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the value.
• For the same operation, directly use the pressure regulating handle to apply pressure clockwise to the verification points of 30MPa, 40MPa, 50MPa, and 60MPa. Once the pressure stabilizes, tap the verified pressure gauge and record the values.
2. Decompression process
Directly use the pressure regulating handle to decrease pressure counterclockwise, reducing it to the verification points of 50MPa, 40MPa, 30MPa, 20MPa, and 10MPa. Check the pressure values of the verified pressure gauge, tap the verified pressure gauge, and record the indicated value after it stabilizes.
3. Pressure relief process
Open the pressure relief valve, gradually release the pressure inside the pipeline, and when the pressure relief reaches 0, the measurement is complete.
III. Precautions
Standardized use
• The pressure transmitting medium needs to be clean and free of impurities to prevent blockage of pipelines, damage to the hydraulic platform, and pressure gauge.
• Operations should follow standard procedures. The ALKT9801 hydraulic calibration bench should be pressurized and depressurized slowly to avoid pressure shock.
• Increase the pressure to the verification point and record the indicated value after the standard gauge's reading stabilizes. For the verified pressure gauge, gently tap the gauge case after the pressure stabilizes, and take the indicated value after tapping as the accurate reading to ensure accurate recording.
• After the inspection is completed, slowly release the pressure to zero and ensure there is no residual pressure before removing the instrument being inspected.
Maintenance and calibration cycle
• Regularly conduct metrological verification on the Precision digital pressure gauge alkc400D (it is recommended to do so once a year) to ensure that the measurement accuracy meets the required standards.
• When the pressure gauge is not in use, the power should be turned off. If it is not used for a long time, it should be charged once every three months.
• If the hydraulic bench is not used for a long time, it needs to be depressurized before storage. Pour out the liquid in the reservoir cup and cover it with a cloth cover to prevent dust from entering the instrument.
IV. Conclusion
This calibration utilizes the AILEIKE ALKT9801 hydraulic calibration bench paired with the ALKC400D precision digital pressure gauge to conduct a full-range calibration test for the Y-100BF pointer pressure gauge, covering a range of 0~60MPa. The operation avoids the impacts of vibration, shock, and corrosive media, standardizes pipeline assembly, medium filling, and exhaust operations, performs graded pressure increase and stabilization, returns to lower pressure for reading indication, conducts error evaluation, and verifies basic error and return difference indicators. Standardized calibration procedures and instrument maintenance management are implemented to ensure accurate and complete calibration data, providing reliable metrological support for compliant operations.